- Overview
- Product Parameters
- Company Profile
Basic Info.
Model NO.
HCTE0027
Humidifier
Apply The External Humidification Mod
Refrigeration Method
Water Cooled
Transport Package
Plywood
Specification
SUS304 stainless steel plate
Trademark
H&H
Origin
China
Production Capacity
10 Sets Per Month
Product Description
Walk-In Constant Temperature And Constant Humidity / High And Low Temperature Laboratory
Gb10589-89 Low temperature test box technical conditions;
Gb10592-89 High and low temperature test box technical conditions;
Gb10586-89 Humidity and heat test chamber technical conditions;
Gb11158-89 High temperature test box technical conditions;
Gb2423.1-89 Basic test specification for electrical and electronic products Test A: Low temperature test method;
Gb2423.2-89 Basic test specification for electrical and electronic products Test B: High temperature test method;
Gb2423.3-91 Basic test procedures for electrical and electronic products test Ca: Constant damp heat test method;
Gb2423.4-89 Basic test specification for electrical and electronic products Test Db: Alternating damp heat test method;
Gb2423.22-87 Basic test specification for electrical and electronic products Test N: Temperature change test method;
Gb2424.1-89 Basic environmental testing procedures for electrical and electronic products Test Db: Guidelines for high and low temperature tests;
GB/T5170.2-1996 Verification method of basic parameters of environmental testing equipment for electrical and electronic products.Temperature test equipment.
GB/T5170.5-1996 Verification method of basic parameters of environmental testing equipment for electrical and electronic products.Humid heat test equipment.
Installation Environment:
1.Temperature: 0ºC ~35ºC, relative humidity: 85%, air pressure: 86kPA~106kPa
2.The installation site shall be flat, well ventilated and no strong vibration around the equipment.
3.There is no strong electromagnetic field influence around the equipment, no flammable, explosive, corrosive substances and dust around the equipment.
4.Proper use and maintenance space remains around the equipment.
Standard volume (m3) | 8 | 12 | 20 | 50 |
Internal chamber size (W * H * D) m | 2×2×2 | 3×2×2 | 4×4×5 | 5×2×5 |
Temperature range | Maximum test temperature: 80ºC, 120ºC; minimum test temperature: -60ºC, -40ºC, -20ºC, 0ºC | |||
Humidity range | 30%~98%RH | |||
Temperature deviation | <±2ºC | |||
Temperature uniformity | ±2ºC | |||
Temperature fluctuation | ±0.5ºC | |||
Humidity deviation | ±5.0%RH | |||
Heating rate | 2.5 ° C/min (non-linear, no-load) | |||
Cooling rate | The average value was 1.0ºC / min. | |||
Inner box material | SUS304 stainless steel plate | |||
Outer box material | A3 Cold-rolled steel sheet electrostatic double-sided injection molding / SUS304 stainless steel. | |||
Heat insulation material | Hard matter polyurethane foaming. | |||
Refrigeration method | Water cooled | |||
Cryogen | R404a,R23 | |||
Temperature and humidity sensor | Class A PT100 Platinum resistance. | |||
Heater | Nickel-chromium alloy electric heating type heater. | |||
Humidifier | Apply the external humidification mode. | |||
Supply source | 380V±10%V;50Hz; | |||
Protection system | Power supply phase deficiency and phase sequence error protection, water shortage automatic alarm protection, leakage protection, heater overtemperature protection, heater short-circuit and other overcurrent protection. | |||
Model | HCTE0027 |